Key takeaways
- HTM 01-05 is the national guidance on decontamination in primary care dental practices in England; compliance is seen as evidence of valid quality assurance.
- Essential quality requirements are the baseline every practice must meet; best practice adds measures such as a dedicated decontamination room, and practices should plan how to move towards it.
- Wrapped instruments may be stored for up to a year; unwrapped instruments for a day in the clinical area or a week in a non-clinical area.
- Audit decontamination at least every six months and keep audit records on the premises for at least two years.
- Equipment must be validated, maintained and tested, and staff must have training records and hepatitis B evidence.
What HTM 01-05 is and why it matters
Health Technical Memorandum 01-05 sets out how primary care dental practices in England should decontaminate reusable instruments. It was first published in 2009, and the current text is the 2013 edition[1]. It follows the essential principles of the Health and Social Care Act 2008 code of practice on the prevention and control of infections[1], which applies to providers of primary dental care[2].
The guidance matters because inspectors use it. HTM 01-05 says compliance will be seen as indicative of the presence of valid quality assurance systems[1]. In practice, a decontamination room or area that follows the guidance, with complete records, is one of the clearest pieces of evidence a dental practice can show that care is safe.
The volume of work is large. NHS dentists in England delivered 76.8 million units of dental activity in 2025/26, 5.1% more than the year before[3], and every examination, filling or extraction relies on instruments that have passed through the decontamination cycle.
Essential quality requirements and best practice
The guidance works on two levels. Essential quality requirements are the baseline that shows a dental service is operating safely and responsibly with respect to decontamination. Best practice goes further, for example by separating decontamination from the clinical treatment area in a dedicated room. HTM 01-05 says practices should carry out an assessment of the changes needed to move from essential quality to best practice[1], and that where new practices or premises are planned, the full best practice provisions should be used wherever reasonably practicable[1].
| Area | Essential quality (baseline) | Moving towards best practice |
|---|---|---|
| Where decontamination happens | Clear dirty-to-clean flow, which may be within the surgery | Dedicated decontamination room or rooms used only for that purpose |
| Cleaning | Validated manual or automated cleaning with inspection | Validated washer-disinfector as the main cleaning method |
| Storage | Dry, protected storage within set time limits | Storage away from clinical contamination with controlled rotation |
| Hand hygiene | Hand-wash facilities for decontamination staff | Separate wash-hand basins dedicated to decontamination |
| Quality system | Policies, logs, six-monthly audit | Documented plan and timetable for best practice |
The table summarises themes, not the full text. Use the guidance itself for the detail, and record your own assessment and plan so an inspector can see where you are and where you are heading.
The decontamination cycle step by step
Transfer
Move used instruments from the surgery to the decontamination area in a covered, rigid container, separating sharps and single-use items at the chairside.
Clean
Clean instruments using a validated process, preferably a washer-disinfector. Where manual cleaning is used, follow the sink arrangement the guidance describes and protect staff from splashing.
Inspect
Check every instrument under magnification and good light for residual debris and damage, and re-clean or discard as needed.
Sterilise
Sterilise in a validated steam steriliser. The guidance's checklist asks whether steam sterilisation at 134 to 138°C for a minimum of three minutes is appropriate for the instruments[1].
Store
Store instruments dry and protected from contamination, wrapped or unwrapped, within the time limits set out below.
Record
Log each cycle, the steriliser test results and any failures, with the operator's name.
Keep a laminated one-page version of your local decontamination procedure on the wall of the decontamination area, with a photo of the dirty-to-clean flow. New staff learn faster, and inspectors can see at a glance that the process is defined.
Storage limits for reprocessed instruments
Storage is one of the most specific parts of the guidance. Wrapped instruments may be stored for up to one year, and practices should develop a quality system so that storage of wrapped instruments does not exceed that[1]. Unwrapped instruments in the clinical area have a maximum storage time of one day, and should be reprocessed at the end of the working day whether or not they were used, or at the start of the next working day; unwrapped instruments in a non-clinical area have a maximum of one week[1].
- Date every wrapped pouch or tray and rotate stock so the oldest is used first.
- Keep unwrapped instruments in covered drawer inserts or mini-racks in cupboards, never on open work surfaces.
- Check storage dates weekly and reprocess anything past its limit.
- Review whether rarely used instruments need to be stored at all, or could be single-use.
Equipment validation, testing and maintenance
HTM 01-05 says equipment used to decontaminate dental instruments should be fit for purpose and validated, which means it should be commissioned, maintained and periodically tested[1]. Inspectors will usually ask to see the test records for each machine.
- Keep a file for each steriliser, washer-disinfector and ultrasonic cleaner with commissioning, validation and service records.
- Record the daily, weekly and periodic tests the manufacturer and guidance specify, signed by the operator.
- Set reminders for annual validation and pressure vessel inspections for sterilisers.
- Record what happens when a test fails: the machine taken out of use, instruments recalled where needed and the repair.
Audit and records
At a minimum, practices should audit their decontamination processes every six months, with an appropriate review depending on the results; the Infection Prevention Society and Department of Health audit tool is strongly recommended[1]. Audit documents should be kept for at least two years, and should not be removed from the premises or destroyed[1].
- Diary the audit for the same two months each year, so it is never late.
- Complete it with the decontamination lead and one other team member, to share knowledge.
- Record every action arising, with an owner and a date.
- Re-check the actions at the next audit and note what changed.
- Keep at least the last four audits together in the decontamination file.
Staff training and protection
A documented training protocol should be in operation, with individual training records for all staff engaged in decontamination[1]. Staff involved in decontamination should also demonstrate current immunisation for hepatitis B and, subject to local policy, tetanus, and must be told the benefits and drawbacks of vaccination[1].
- Train every new dental nurse in your local procedure before they decontaminate unsupervised.
- Refresh training annually and after any change of equipment.
- Keep immunisation evidence in the staff file and review it at appraisal.
- Make sure protective equipment is available in the decontamination area at all times.
Why reception should care
Decontamination capacity is a hidden limit on the diary. If a practice owns a fixed number of instrument sets for a procedure and the cycle takes a set time, booking more of that procedure than the sets allow forces either delays or shortcuts. Ask the decontamination lead how many of each kind of procedure can safely be booked per session, and build that limit into the booking rules.
The same applies to unplanned closures. If a steriliser fails, reception needs a clear script for rebooking patients and a list of what can still go ahead using single-use instruments or stock that is already processed and in date.
Demand makes this more than a theoretical risk. NHS dental activity is rising again, with 37.6 million NHS courses of treatment delivered in England in 2025/26[4], and many practices are adding private sessions on top. When the diary grows faster than instrument stock, the pressure lands on the decontamination area first. Review instrument numbers whenever you add sessions, a clinician or a new treatment type, and budget for extra sets before the bookings arrive rather than after a near miss.
Example: preparing for an inspection
Example (fictional): a three-surgery practice that decontaminates within each surgery decides to review its position before an expected inspection. The decontamination lead finds that the last two six-monthly audits are complete, but that wrapped pouches in one surgery carry no dates and one ultrasonic cleaner has no recorded periodic tests for three months.
The practice dates and rotates all stock, takes the ultrasonic cleaner out of use until it is tested, and writes a short plan with costs for converting a storeroom into a dedicated decontamination room within two years. At inspection, the lead can show both the fixes and the plan to move towards best practice, which is the kind of evidence the guidance asks for.
Frequently asked questions
Is HTM 01-05 mandatory?
It is guidance, but inspectors use it to judge decontamination, and it says compliance is indicative of valid quality assurance systems. Practices are expected to meet the essential quality requirements and plan towards best practice.
How long can sterilised dental instruments be stored?
Wrapped instruments up to one year; unwrapped instruments up to one day in the clinical area or one week in a non-clinical area, under HTM 01-05.
How often should we audit decontamination?
At least every six months, ideally with the Infection Prevention Society audit tool, and keep audit documents on the premises for at least two years.
Do we need a separate decontamination room?
Not to meet essential quality requirements, but a dedicated room is part of best practice. Assess the changes needed and, for new premises, use the best practice provisions wherever reasonably practicable.
References
Department of Health (2013). Health Technical Memorandum 01-05: Decontamination in primary care dental practices (2013 edition) (opens in a new tab) ↑
GOV.UK (Department of Health and Social Care) (2022). Health and Social Care Act 2008: code of practice on the prevention and control of infections and related guidance (opens in a new tab) ↑
NHS Business Services Authority (2026). Dental statistics - England 2025/26 (opens in a new tab) ↑
Care Quality Commission (2026). State of Care 2025/26: dental (opens in a new tab) ↑